Frequently Used Consumables in Flow Cytometry and Cell Sorting Experiments

Items Purpose Example 
Cell strainer To remove larger debris from the sample. This is an essential and mandatory step to protect the instruments from clogging.40 um Flowmi (VWR: 10032-802)
35 um cell strainer with 5ml FACS tube (FALCON 352235, VWR: CA210080-948)
35 um cell strainer cap only (This can fit 12x75mm 5ml tube) 
FACS buffer To protect cells from stress.  To reduce non-specific binding of antibodies eBioScience FC staining buffer 
Biolegend Cell Staining buffer 
In-house (PBS/HBBS + BSA and/or FBS) 
Tubes To load samples to the cytometer or cell sorter Sterile PP tube with cap is recommended for receiving sorted cells. 12×75 mm tube (5 mL tube, FACS tube)
Plain, no cap, non-sterile (352002, VWR: 60819-794) 
Plain, sterile with cap (352063, VWR: 60819-728) 

Discussion: at which step do we filter cells?

Step 2: Cell Strainer

Pore SizeGood For
100 umInitial filtration to remove larger debris
70 umInitial filtration to remove larger debris
40 umFinal filtration or filtration for single cell suspension
VMR: 104027-370
Flowmi cell strainer
Flowmi cell strainer (70um, 40um available from VMR)

Good for small volume

VMR 10032-802

Note: 

96-well plate can be used for delivering samples to CytoFLEX LX cell analyzer. We recommend U or V bottom shaped plates. The bottom size of the plate has to be (8.5 cm x 12.7 cm) to fit the loader. 

If you plan to perform sample prep (staining) in the plate, pay close attention to 1) plate material/coating, 2) bottom shape, 3) liquid volume. Those factors impact largely to your staining result and cell recovery rate.